- F A C E T -

Laser Angle Measuring Unit
Random Tests in the Laboratory and Production

FACET - click to enlarge

Measuring Parameters

  • facet angle
  • rotation angle
  • back bevel angle
  • symmetry
  • accuracy +/- 1 degree

The measuring is touchless and quick, but without any modification at the cannulae. Thus it is applicable during the grinding process, the assembly process and finally at the finished product.


Technical Performance:

  • HeNe-Laser 0,5 mW
  • power supply unit
  • vertical adjustable laser support
  • cannula support and fixation
  • horizontal screen
    measuring scale 60 - 180 degree
    symmetry-scale +/- 40 degree
  • vertical screen
    measuring scale 60 - 180 degree
    symmetry-scale +/- 40 degree
FACET - click to enlarge

Also touchless and 100% in-line works the rotary alignment of the bevel regarding an application on assembly lines, DREHFIX.